ISO 13485 Certification Documentation Assistant
Overview
This skill helps medical device manufacturers prepare comprehensive documentation for ISO 13485:2016 certification. It provides tools, templates, references, and guidance to create, review, and gap-analyze all required Quality Management System (QMS) documentation.
What this skill provides:
- Gap analysis of existing documentation
- Templates for all mandatory documents
- Comprehensive requirements guidance
- Step-by-step documentation creation
- Identification of missing documentation
- Compliance checklists
When to use this skill:
- Starting ISO 13485 certification process
- Conducting gap analysis against ISO 13485
- Creating or updating QMS documentation
- Preparing for certification audit
- Transitioning from FDA QSR to QMSR
- Harmonizing with EU MDR requirements
Core Workflow
1. Assess Current State (Gap Analysis)
When to start here: User has existing documentation and needs to identify gaps
Process:
-
Collect existing documentation:
- Ask user to provide directory of current QMS documents
- Documents can be in any format (.txt, .md, .doc, .docx, .pdf)
- Include any procedures, manuals, work instructions, forms
-
Run gap analysis script:
python scripts/gap_analyzer.py --docs-dir <path_to_docs> --output gap-report.json -
Review results:
- Identify which of the 31 required procedures are present
- Identify missing key documents (Quality Manual, MDF, etc.)
- Calculate compliance percentage
- Prioritize missing documentation
-
Present findings to user:
- Summarize what exists
- Clearly list what's missing
- Provide prioritized action plan
- Estimate effort required
Output: Comprehensive gap analysis report with prioritized action items
2. Understand Requirements (Reference Consultation)
When to use: User needs to understand specific ISO 13485 requirements
Available references:
references/iso-13485-requirements.md- Complete clause-by-clause breakdownreferences/mandatory-documents.md- All 31 required procedures explainedreferences/gap-analysis-checklist.md- Detailed compliance checklistreferences/quality-manual-guide.md- How to create Quality Manual
How to use:
-
For specific clause questions:
- Read relevant section from
iso-13485-requirements.md - Explain requirements in plain language
- Provide practical examples
- Read relevant section from
-
For document requirements:
- Consult
mandatory-documents.md - Explain what must be documented
- Clarify when documents are applicable vs. excludable
- Consult
-
For implementation guidance:
- Use
quality-manual-guide.mdfor policy-level documents - Provide step-by-step creation process
- Show examples of good vs. poor implementation
- Use
Key reference sections to know:
- Clause 4: QMS requirements, documentation, risk management, software validation
- Clause 5: Management responsibility, quality policy, objectives, management review
- Clause 6: Resources, competence, training, infrastructure
- Clause 7: Product realization, design, purchasing, production, traceability
- Clause 8: Measurement, audits, CAPA, complaints, data analysis
3. Create Documentation (Template-Based Generation)
When to use: User needs to create specific QMS documents
Available templates:
- Quality Manual:
assets/templates/quality-manual-template.md - CAPA Procedure:
assets/templates/procedures/CAPA-procedure-template.md - Document Control:
assets/templates/procedures/document-control-procedure-template.md
Process for document creation:
-
Identify what needs to be created:
- Based on gap analysis or user request
- Prioritize critical documents first (Quality Manual, CAPA, Complaints, Audits)
-
Select appropriate template:
- Use Quality Manual template for QM
- Use procedure templates as examples for SOPs
- Adapt structure to organization's needs
-
Customize template with user-specific information:
- Replace all placeholder text: [COMPANY NAME], [DATE], [NAME], etc.
- Tailor scope to user's actual operations
- Add or remove sections based on applicability
- Ensure consistency with organization's processes
-
Key customization areas:
- Company information and addresses
- Product types and classifications
- Applicable regulatory requirements
- Organization structure and responsibilities
- Actual processes and procedures
- Document numbering schemes
- Exclusions and justifications
-
Validate completeness:
- All required sections present
- All placeholders replaced
- Cross-references correct
- Approval sections complete
Document creation priority order:
Phase 1 - Foundation (Critical):
- Quality Manual
- Quality Policy and Objectives
- Document Control procedure
- Record Control procedure
Phase 2 - Core Processes (High Priority): 5. Corrective and Preventive Action (CAPA) 6. Complaint Handling 7. Internal Audit 8. Management Review 9. Risk Management
Phase 3 - Product Realization (High Priority): 10. Design and Development (if applicable) 11. Purchasing 12. Production and Service Provision 13. Control of Nonconforming Product
Phase 4 - Supporting Processes (Medium Priority): 14. Training and Competence 15. Calibration/Control of M&M Equipment 16. Process Validation 17. Product Identification and Traceability
Phase 5 - Additional Requirements (Medium Priority): 18. Feedback and Post-Market Surveillance 19. Regulatory Reporting 20. Customer Communication 21. Data Analysis
Phase 6 - Specialized (If Applicable): 22. Installation (if applicable) 23. Servicing (if applicable) 24. Sterilization (if applicable) 25. Contamination Control (if applicable)
4. Develop Specific Documents
Creating a Quality Manual
Process:
-
Read the comprehensive guide:
- Read
references/quality-manual-guide.mdin full - Understand structure and required content
- Review examples provided
- Read
-
Gather organization information:
- Legal company name and addresses
- Product types and classifications
- Organizational structure
- Applicable regulations
- Scope of operations
- Any exclusions needed
-
Use template:
- Start with
assets/templates/quality-manual-template.md - Follow structure exactly (required by ISO 13485)
- Replace all placeholders
- Start with
-
Complete required sections:
- Section 0: Document control, approvals
- Section 1: Introduction, company overview
- Section 2: Scope and exclusions (critical - must justify exclusions)
- Section 3: Quality Policy (must be signed by top management)
- Sections 4-8: Address each ISO 13485 clause at policy level
- Appendices: Procedure list, org chart, process map, definitions
-
Key requirements:
- Must reference all 31 documented procedures (Appendix A)
- Must describe process interactions (Appendix C - create process map)
- Must define documentation structure (Section 4.2)
- Must justify any exclusions (Section 2.4)
-
Validation checklist:
- All required content per ISO 13485 Clause 4.2.2
- Quality Policy signed by top management
- All exclusions justified
- All procedures listed in Appendix A
- Process map included
- Organization chart included
Creating Procedures (SOPs)
General approach for all procedures:
-
Understand the requirement:
- Read relevant clause in
references/iso-13485-requirements.md - Understand WHAT must be documented
- Identify WHO, WHEN, WHERE for your organization
- Read relevant clause in
-
Use template structure:
- Follow CAPA or Document Control templates as examples
- Standard sections: Purpose, Scope, Definitions, Responsibilities, Procedure, Records, References
- Keep procedures clear and actionable
-
Define responsibilities clearly:
- Identify specific roles (not names)
- Define responsibilities for each role
- Ensure coverage of all required activities
-
Document the "what" not excessive "how":
- Procedures should define WHAT must be done
- Detailed HOW-TO goes in Work Instructions (Tier 3)
- Strike balance between guidance and flexibility
-
Include required elements:
- All elements specified in ISO 13485 clause
- Records that must be maintained
- Responsibilities for each activity
- References to related documents
Example: Creating CAPA Procedure
- Read ISO 13485 Clauses 8.5.2 and 8.5.3 from references
- Use
assets/templates/procedures/CAPA-procedure-template.md - Customize:
- CAPA prioritization criteria for your organization
- Root cause analysis methods you'll use
- Approval authorities and responsibilities
- Timeframes based on your operations
- Integration with complaint handling, audits, etc.
- Add forms as attachments:
- CAPA Request Form
- Root Cause Analysis Worksheet
- Action Plan Template
- Effectiveness Verification Checklist
Creating Medical Device Files (MDF)
What is an MDF:
- File for each medical device type or family
- Replaces separate DHF, DMR, DHR (per FDA QMSR harmonization)
- Contains all documentation about the device
Required contents per ISO 13485 Clause 4.2.3:
- General description and intended use
- Label and instructions for use specifications
- Product specifications
- Manufacturing specifications
- Procedures for purchasing, manufacturing, servicing
- Procedures for measuring and monitoring
- Installation requirements (if applicable)
- Risk management file(s)
- Verification and validation information
- Design and development file(s) (when applicable)
Process:
- Identify each device type or family
- Create MDF structure (folder or binder)
- Collect or create each required element
- Ensure traceability between documents
- Maintain as living document (update with changes)
5. Conduct Comprehensive Gap Analysis
When to use: User wants detailed assessment of all requirements
Process:
-
Use comprehensive checklist:
- Open
references/gap-analysis-checklist.md - Work through clause by clause
- Mark status for each requirement: Compliant, Partial, Non-compliant, N/A
- Open
-
For each clause:
- Read requirement description
- Identify existing evidence
- Note gaps or deficiencies
- Define action required
- Assign responsibility and target date
-
Summarize by clause:
- Calculate compliance percentage per clause
- Identify highest-risk gaps
- Prioritize actions
-
Create action plan:
- List all gaps
- Prioritize: Critical > High > Medium > Low
- Assign owners and dates
- Estimate resources needed
-
Output:
- Completed gap analysis checklist
- Summary report with compliance percentages
- Prioritized action plan
- Timeline and milestones
Common Scenarios
Scenario 1: Starting from Scratch
User request: "We're a medical device startup and need to implement ISO 13485. Where do we start?"
Approach:
-
Explain the journey:
- ISO 13485 requires comprehensive QMS documentation
- Typically 6-12 months for full implementation
- Can be done incrementally
-
Start with foundation:
- Quality Policy and Objectives
- Quality Manual
- Organization structure and responsibilities
-
Follow the priority order:
- Use Phase 1-6 priority list above
- Create documents in logical sequence
- Build on previously created documents
-
Key milestones:
- Month 1-2: Foundation documents (Quality Manual, policies)
- Month 3-4: Core processes (CAPA, Complaints, Audits)
- Month 5-6: Product realization processes
- Month 7-8: Supporting processes
- Month 9-10: Internal audits and refinement
- Month 11-12: Management review and certification audit
Scenario 2: Gap Analysis for Existing QMS
User request: "We have some procedures but don't know what we're missing for ISO 13485."
Approach:
-
Run automated gap analysis:
- Ask for document directory
- Run
scripts/gap_analyzer.py - Review automated findings
-
Conduct detailed assessment:
- Use comprehensive checklist for user's specific situation
- Go deeper than automated analysis
- Assess quality of existing documents, not just presence
-
Provide prioritized gap list:
- Missing mandatory procedures
- Incomplete procedures
- Quality issues with existing documents
- Missing records or forms
-
Create remediation plan:
- High priority: Safety-related, regulatory-required
- Medium priority: Core QMS processes
- Low priority: Improvement opportunities
Scenario 3: Creating Specific Document
User request: "Help me create a CAPA procedure."
Approach:
-
Explain requirements:
- Read ISO 13485 Clauses 8.5.2 and 8.5.3 from references
- Explain what must be in CAPA procedure
- Provide examples of good CAPA processes
-
Use template:
- Start with CAPA procedure template
- Explain each section's purpose
- Show what needs customization
-
Gather user-specific info:
- How are CAPAs initiated in their organization?
- Who are the responsible parties?
- What prioritization criteria make sense?
- What RCA methods will they use?
- What are appropriate timeframes?
-
Create customized procedure:
- Replace all placeholders
- Adapt to user's processes
- Ensure completeness
-
Add supporting materials:
- CAPA request form
- RCA worksheets
- Action plan template
- Effectiveness verification checklist
Scenario 4: Updating for Regulatory Changes
User request: "We need to update our QMS for FDA QMSR harmonization."
Approach:
-
Explain changes:
- FDA 21 CFR Part 820 harmonized with ISO 13485
- Now called QMSR (effective Feb 2, 2026)
- Key change: Medical Device File replaces DHF/DMR/DHR
-
Review current documentation:
- Identify documents referencing QSR
- Find separate DHF, DMR, DHR structures
- Check for ISO 13485 compliance gaps
-
Update strategy:
- Update references from QSR to QMSR
- Consolidate DHF/DMR/DHR into Medical Device Files
- Add any missing ISO 13485 requirements
- Maintain backward compatibility during transition
-
Create transition plan:
- Update Quality Manual
- Update MDF procedure
- Reorganize device history files
- Train personnel on changes
Scenario 5: Preparing for Certification Audit
User request: "We have our documentation ready. How do we prepare for the certification audit?"
Approach:
-
Conduct readiness assessment:
- Use comprehensive gap analysis checklist
- Review all documentation for completeness
- Verify records exist for all required items
- Check for consistent implementation
-
Pre-audit checklist:
- All 31 procedures documented and approved
- Quality Manual complete with all required content
- Medical Device Files complete for all products
- Internal audit completed with findings addressed
- Management review completed
- Personnel trained on QMS procedures
- Records maintained per retention requirements
- CAPA system functional with effectiveness demonstrated
- Complaints system operational
-
Conduct mock audit:
- Use ISO 13485 requirements as audit criteria
- Sample records to verify consistent implementation
- Interview personnel to verify understanding
- Identify any non-conformances
-
Address findings:
- Correct any deficiencies
- Document corrections
- Verify effectiveness
-
Final preparation:
- Brief management and staff
- Prepare audit schedule
- Organize evidence and records
- Designate escorts and support personnel
Best Practices
Document Development
-
Start at policy level, then add detail:
- Quality Manual = policy level
- Procedures = what, who, when
- Work Instructions = detailed how-to
- Forms = data collection
-
Maintain consistency:
- Use same terminology throughout
- Cross-reference related documents
- Keep numbering scheme consistent
- Update all related documents together
-
Write for your audience:
- Clear, simple language
- Avoid jargon
- Define technical terms
- Provide examples where helpful
-
Make procedures usable:
- Action-oriented language
- Logical flow
- Clear responsibilities
- Realistic timeframes
Exclusions
When you can exclude:
- Design and development (if contract manufacturer only)
- Installation (if product requires no installation)
- Servicing (if not offered)
- Sterilization (if non-sterile product)
Justification requirements:
- Must be in Quality Manual
- Must explain why excluded
- Cannot exclude if process performed
- Cannot affect ability to provide safe, effective devices
Example good justification:
"Clause 7.3 Design and Development is excluded. ABC Company operates as a contract manufacturer and produces medical devices according to complete design specifications provided by customers. All design activities are performed by the customer and ABC Company has no responsibility for design inputs, outputs, verification, validation, or design changes."
Example poor justification:
"We don't do design." (Too brief, doesn't explain why or demonstrate no impact)
Common Mistakes to Avoid
-
Copying ISO 13485 text verbatim
- Write in your own words
- Describe YOUR processes
- Make it actionable for your organization
-
Making procedures too detailed
- Procedures should be stable
- Excessive detail belongs in work instructions
- Balance guidance with flexibility
-
Creating documents in isolation
- Ensure consistency across QMS
- Cross-reference related documents
- Build on previously created documents
-
Forgetting records
- Every procedure should specify records
- Define retention requirements
- Ensure records actually maintained
-
Inadequate approval
- Quality Manual must be signed by top management
- All procedures must be properly approved
- Train staff before documents become effective
Resources
scripts/
gap_analyzer.py- Automated tool to analyze existing documentation and identify gaps against ISO 13485 requirements
references/
iso-13485-requirements.md- Complete breakdown of ISO 13485:2016 requirements clause by clausemandatory-documents.md- Detailed list of all 31 required procedures plus other mandatory documentsgap-analysis-checklist.md- Comprehensive checklist for detailed gap assessmentquality-manual-guide.md- Step-by-step guide for creating a compliant Quality Manual
assets/templates/
quality-manual-template.md- Complete template for Quality Manual with all required sectionsprocedures/CAPA-procedure-template.md- Example CAPA procedure following best practicesprocedures/document-control-procedure-template.md- Example document control procedure
Quick Reference
The 31 Required Documented Procedures
- Risk Management (4.1.5)
- Software Validation (4.1.6)
- Control of Documents (4.2.4)
- Control of Records (4.2.5)
- Internal Communication (5.5.3)
- Management Review (5.6.1)
- Human Resources/Competence (6.2)
- Infrastructure Maintenance (6.3) - when applicable
- Contamination Control (6.4.2) - when applicable
- Customer Communication (7.2.3)
- Design and Development (7.3.1-10) - when applicable
- Purchasing (7.4.1)
- Verification of Purchased Product (7.4.3)
- Production Control (7.5.1)
- Product Cleanliness (7.5.2) - when applicable
- Installation (7.5.3) - when applicable
- Servicing (7.5.4) - when applicable
- Process Validation (7.5.6) - when applicable
- Sterilization Validation (7.5.7) - when applicable
- Product Identification (7.5.8)
- Traceability (7.5.9)
- Customer Property (7.5.10) - when applicable
- Preservation of Product (7.5.11)
- Control of M&M Equipment (7.6)
- Feedback (8.2.1)
- Complaint Handling (8.2.2)
- Regulatory Reporting (8.2.3)
- Internal Audit (8.2.4)
- Process Monitoring (8.2.5)
- Product Monitoring (8.2.6)
- Control of Nonconforming Product (8.3)
- Corrective Action (8.5.2)
- Preventive Action (8.5.3)
(Note: Traditional count is "31 procedures" though list shows more because some are conditional)
Key Regulatory Requirements
FDA (United States):
- 21 CFR Part 820 (now QMSR) - harmonized with ISO 13485 as of Feb 2026
- Device classification determines requirements
- Establishment registration and device listing required
EU (European Union):
- MDR 2017/745 (Medical Devices Regulation)
- IVDR 2017/746 (In Vitro Diagnostic Regulation)
- Technical documentation requirements
- CE marking requirements
Canada:
- Canadian Medical Devices Regulations (SOR/98-282)
- Device classification system
- Medical Device Establishment License (MDEL)
Other Regions:
- Australia TGA, Japan PMDA, China NMPA, etc.
- Often require or recognize ISO 13485 certification
Document Retention
Minimum retention: Lifetime of medical device as defined by organization
Typical retention periods:
- Design documents: Life of device + 5-10 years
- Manufacturing records: Life of device
- Complaint records: Life of device + 5-10 years
- CAPA records: 5-10 years minimum
- Calibration records: Retention period of equipment + 1 calibration cycle
Always comply with applicable regulatory requirements which may specify longer periods.
Getting Started
First-time users should:
- Read
references/iso-13485-requirements.mdto understand the standard - If you have existing documentation, run gap analysis script
- Create Quality Manual using template and guide
- Develop procedures in priority order
- Use comprehensive checklist for final validation
For specific tasks:
- Creating Quality Manual → See Section 4 and use quality-manual-guide.md
- Creating CAPA procedure → See Section 4 and use CAPA template
- Gap analysis → See Section 1 and 5
- Understanding requirements → See Section 2
Need help? Start by describing your situation: what stage you're at, what you have, and what you need to create.
GitHub 저장소
연관 스킬
content-collections
메타이 스킬은 콘텐츠 콜렉션(Content Collections)을 위한 프로덕션 검증된 설정을 제공합니다. 콘텐츠 콜렉션은 Markdown/MDX 파일을 Zod 검증이 포함된 타입 안전한 데이터 콜렉션으로 변환해주는 TypeScript 최우선 도구입니다. 블로그, 문서 사이트 또는 콘텐츠 중심의 Vite + React 애플리케이션을 구축할 때 타입 안전성과 자동 콘텐츠 검증을 보장하기 위해 사용하세요. Vite 플러그인 구성과 MDX 컴파일부터 배포 최적화 및 스키마 검증에 이르기까지 모든 것을 다룹니다.
polymarket
메타이 스킬은 개발자들이 Polymarket 예측 시장 플랫폼을 활용한 애플리케이션을 구축할 수 있도록 지원하며, 거래 및 시장 데이터를 위한 API 통합 기능을 포함합니다. 또한 WebSocket을 통한 실시간 데이터 스트리밍을 제공하여 실시간 거래와 시장 활동을 모니터링할 수 있습니다. 이를 통해 거래 전략을 구현하거나 실시간 시장 업데이트를 처리하는 도구를 생성하는 데 활용할 수 있습니다.
creating-opencode-plugins
메타이 스킬은 개발자들이 명령어, 파일, LSP 작업 등 25개 이상의 이벤트 유형에 연결되는 OpenCode 플러그인을 만들 수 있도록 돕습니다. JavaScript/TypeScript 모듈을 위한 플러그인 구조, 이벤트 API 명세, 구현 패턴을 제공합니다. OpenCode AI 어시스턴트의 라이프사이클을 사용자 정의 이벤트 기반 로직으로 가로채거나, 모니터링하거나, 확장해야 할 때 사용하세요.
sglang
메타SGLang은 RadixAttention 프리픽스 캐싱을 활용하여 JSON, 정규식, 에이전트 워크플로우를 위한 고속 구조화 생성에 특화된 고성능 LLM 서빙 프레임워크입니다. 특히 반복되는 프리픽스가 있는 작업에서 상당히 빠른 추론 속도를 제공하여 복잡한 구조화 출력 및 다중 턴 대화에 이상적입니다. 제약 디코딩이 필요하거나 광범위한 프리픽스 공유가 있는 애플리케이션을 구축할 때는 vLLM과 같은 대안보다 SGLang을 선택하십시오.