This skill analyzes GC/HPLC chromatograms to verify system suitability, identify and integrate peaks, and calculate chromatographic metrics. It enables automated quality assessment for reliable quantitation in analytical workflows. Developers can use it to programmatically process chromatographic data and generate standardized reports.

快速安装

Claude Code

推荐
主要方式
npx skills add pjt222/agent-almanac -a claude-code
插件命令备选方式
/plugin add https://github.com/pjt222/agent-almanac
Git 克隆备选方式
git clone https://github.com/pjt222/agent-almanac.git ~/.claude/skills/interpret-chromatogram

在 Claude Code 中复制并粘贴此命令以安装该技能

技能文档

Interpret a Chromatogram

Systematic interpretation of GC and HPLC chromatograms covering system suitability verification, peak identification, integration, calculation of chromatographic parameters, and assessment of peak quality for confident qualitative and quantitative results.

When to Use

  • Reviewing chromatographic data before reporting analytical results
  • Verifying that a system suitability test passes before running a sample sequence
  • Identifying unknown peaks or confirming known analytes by retention time or spectral data
  • Troubleshooting unexpected peaks, baseline anomalies, or integration artifacts
  • Training analysts on chromatographic data interpretation

Inputs

Required

  • Chromatogram data: Digital or printed chromatogram with time axis and detector response axis
  • Reference standard data: Retention times and responses of known analytes under the same method conditions
  • Method parameters: Column, mobile phase/carrier gas, temperature/gradient program, detector settings

Optional

  • Spectral data: UV-Vis spectra (DAD), mass spectra (MS), or other spectral information for peak confirmation
  • Previous chromatograms: Historical data from the same method for trend comparison
  • System suitability criteria: Acceptance limits from the method or regulatory standard
  • Sample preparation details: Dilution factors, extraction recovery, internal standard concentration

Procedure

Step 1: Verify System Suitability

Confirm that the chromatographic system is performing within specification before interpreting sample data.

ParameterTypical SpecificationCalculation
Retention time RSD<= 1.0%RSD of tR over n >= 5 injections
Peak area RSD<= 2.0% (assay), <= 5.0% (impurity)RSD of area over n >= 5 injections
Tailing factor (T)0.8-2.0 (USP), ideally 0.9-1.2T = W0.05 / (2 * f) where W0.05 = width at 5% height, f = front half-width
Resolution (Rs)>= 1.5 (baseline), >= 2.0 (regulated)Rs = 2(tR2 - tR1) / (w1 + w2)
Theoretical plates (N)Per column spec (e.g., >= 2000)N = 16(tR / w)^2 or N = 5.54(tR / w0.5)^2
Capacity factor (k')2.0-10.0 for primary analytek' = (tR - t0) / t0
  1. Locate the system suitability injections (typically 5-6 replicates of a reference standard at the start of the sequence).
  2. Calculate each parameter from the table above.
  3. Compare calculated values against the method's acceptance criteria.
  4. If any parameter fails, the system is not suitable -- do not proceed to sample interpretation until the issue is resolved.
  5. Document all system suitability results in the batch record.

Got: All system suitability parameters within specification, confirming the system is fit for purpose.

If fail: If retention time RSD fails, check for temperature instability, mobile phase preparation errors, or column degradation. If tailing factor fails, inspect the inlet liner (GC) or column frit (HPLC). If resolution fails, check column performance with a dedicated test mix and replace if necessary.

Step 2: Identify Peaks

  1. Compare each peak's retention time (tR) against the reference standard chromatogram.
    • Acceptable retention time match: within +/- 2% of the reference tR (or +/- 0.1 min for short runs).
  2. For ambiguous identifications, use co-injection (spiking): add reference standard to the sample and re-inject. The target peak should increase without broadening or shouldering.
  3. For DAD-equipped HPLC: compare the UV-Vis spectrum of each peak against a spectral library.
    • Spectral match index >= 990 (out of 1000) for positive identification.
    • Check spectral purity across the peak (front, apex, tail spectra should overlay).
  4. For MS-equipped systems: confirm molecular ion (m/z) and key fragment ions against reference spectra.
  5. Flag any peak that cannot be identified -- report it as "unknown" with its retention time and relative response.

Got: All target analytes identified by retention time matching, with spectral confirmation where available. Unknown peaks flagged with retention time and area.

If fail: If retention times have shifted uniformly, a systematic change has occurred (column aging, temperature drift, mobile phase error). Re-inject the reference standard to establish current retention times before re-evaluating.

Step 3: Perform Peak Integration

  1. Select integration mode:
    • Automatic integration with data system defaults as a starting point
    • Manual adjustment only when automatic integration demonstrably misplaces baseline or peak boundaries
  2. Set integration parameters:
    • Baseline detection sensitivity (slope sensitivity / threshold)
    • Minimum peak area or height to reject noise
    • Peak width parameter matching the narrowest expected peak
  3. Verify baseline placement:
    • Baseline should connect the start and end of each peak at the true chromatographic baseline
    • For overlapping peaks, use valley-to-valley or perpendicular drop methods as specified by the method
    • For gradient methods, baseline may rise -- use a tangent skim or exponential skim for peaks on a rising baseline
  4. Check for integration errors:
    • Split peaks integrated as two when they should be one
    • Shoulder peaks merged into the main peak when they should be separate
    • Noise spikes integrated as peaks
    • Baseline drawn through a peak (negative peak clipping)
  5. Record the final integration parameters and any manual adjustments with justification in an audit trail.

Got: All target peaks integrated with correct baseline placement, no artifacts included, and all manual adjustments documented with rationale.

If fail: If the automatic integrator consistently mishandles a particular peak shape, create a timed-events integration method with custom parameters for that retention window. Never manually adjust integration to achieve a desired result -- adjustments must be scientifically justified.

Step 4: Calculate Chromatographic Parameters

Calculate the following for all reported peaks:

  1. Resolution (Rs) between adjacent peaks:
    • Rs = 2(tR2 - tR1) / (w1 + w2)
    • Rs >= 1.5 indicates baseline separation; Rs >= 2.0 provides margin for routine use
  2. Tailing factor (T) at 5% peak height:
    • T = W0.05 / (2f)
    • T = 1.0 is perfectly symmetric; T > 2.0 indicates significant tailing
  3. Theoretical plates (N):
    • N = 16(tR / w)^2 using baseline width, or N = 5.54(tR / w0.5)^2 using half-height width
    • Higher N means better column efficiency
  4. Capacity factor (k'):
    • k' = (tR - t0) / t0 where t0 is the dead time (void volume / flow rate)
    • Ideal range: 2-10 for good separation with reasonable run time
  5. Selectivity factor (alpha) between critical pair:
    • alpha = k'2 / k'1
    • alpha > 1.05 is needed for adequate separation
  6. Tabulate results for all analytes and compare against method specifications.

Got: All chromatographic parameters calculated, tabulated, and compared to acceptance criteria. Critical pair resolution and plate count documented.

If fail: If calculated plates are significantly below the column specification, the column may be degraded -- test with a fresh standard and compare to historical data. If parameters drift within a sequence, investigate instrument stability.

Step 5: Assess Peak Quality

  1. Symmetry: Peaks should be Gaussian or near-Gaussian. Significant fronting (T < 0.8) suggests column overload; tailing (T > 1.5) suggests secondary interactions or dead volume.
  2. Baseline separation: For quantitative work, critical pairs must be baseline-resolved. If valley between peaks does not return to baseline, note the percentage valley and assess impact on accuracy.
  3. Peak width consistency: Peaks that are significantly broader than expected (compared to the standard) may indicate on-column degradation, extra-column band broadening, or injection issues.
  4. Spectral purity (DAD/MS): If the purity index indicates spectral inhomogeneity across the peak, a co-eluting impurity is likely present. Consider adjusting the method for better resolution.
  5. Negative peaks or baseline disturbances: Negative peaks in UV indicate the sample solvent absorbs more than the mobile phase at the detection wavelength -- normal for the solvent front but abnormal elsewhere.
  6. Ghost peaks: Peaks present in the blank injection indicate carryover, contaminated mobile phase, or column bleed. Identify the source before reporting sample results.
  7. Summarize overall chromatographic quality and note any limitations on the reported results.

Got: Peak quality assessed for all target analytes; any anomalies (tailing, co-elution, ghost peaks) documented with their potential impact on data quality.

If fail: If significant quality issues are found (co-elution confirmed by spectral impurity, ghost peaks at analyte retention times), the data may not be reportable. Flag the results, investigate root cause, and re-run after corrective action.

Validation

  • System suitability parameters calculated and within specification
  • All target analytes identified by retention time (+/- spectral confirmation)
  • Unknown peaks flagged with retention time and area
  • Integration performed with correct baseline placement; manual adjustments documented
  • Resolution, tailing, plates, and capacity factor calculated for all peaks
  • Peak quality assessed -- no unresolved co-elutions affecting quantitation
  • Ghost peaks and carryover evaluated via blank injection
  • Results tabulated and compared against method acceptance criteria

Pitfalls

  • Accepting automatic integration without review: Data systems can misplace baselines, especially for shoulders, small peaks near large ones, and gradient baselines. Every chromatogram must be visually reviewed.
  • Confusing retention time shift with a new peak: Uniform retention time shifts (all peaks move together) indicate a systematic change, not new compounds. Re-inject the standard to recalibrate before making identification calls.
  • Reporting peaks below the noise level: Peaks with signal-to-noise ratio below 3 (detection) or 10 (quantitation) should not be identified or quantitated. Calculate S/N explicitly for trace-level peaks.
  • Ignoring the solvent front: The void volume peak is not an analyte. Ensure t0 is correctly identified and excluded from analyte reporting.
  • Manual integration to achieve a target result: Adjusting integration to make a result pass specification is data falsification. All integration changes must be scientifically justified and audit-trailed.
  • Neglecting spectral purity checks: A clean-looking peak can hide a co-eluting impurity. Always check peak purity when DAD or MS data is available.

Related Skills

  • develop-gc-method -- method development for the GC technique producing the chromatogram
  • develop-hplc-method -- method development for the HPLC technique producing the chromatogram
  • troubleshoot-separation -- diagnosing problems identified during chromatogram interpretation
  • validate-analytical-method -- formal validation of the method generating the chromatographic data
  • interpret-mass-spectrum -- detailed interpretation of MS data for GC-MS and LC-MS peak confirmation

GitHub 仓库

pjt222/agent-almanac
路径: i18n/caveman-lite/skills/interpret-chromatogram
0
agentsagentskillsai-assisted-developmentclaude-codeskillsteams

相关推荐技能

executing-plans

设计

该Skill用于当开发者提供完整实施计划时,以受控批次方式执行代码实现。它会先审阅计划并提出疑问,然后分批次执行任务(默认每批3个任务),并在批次间暂停等待审查。关键特性包括分批次执行、内置检查点和架构师审查机制,确保复杂系统实现的可控性。

查看技能

requesting-code-review

设计

该Skill可在完成任务、实现主要功能或合并代码前自动调度代码审查子代理,确保实现符合需求和计划。它支持通过指定git SHA范围进行精准的代码变更审查,帮助开发者在关键节点及时发现潜在问题。核心原则是"早审查、勤审查",适用于开发流程的各个关键阶段。

查看技能

connect-mcp-server

设计

这个Skill指导开发者如何将MCP服务器连接到Claude Code,支持HTTP、stdio和SSE三种传输协议。它涵盖了从安装配置到认证安全的完整流程,适用于集成GitHub、Notion、数据库等外部服务。当开发者需要添加集成、配置外部工具或提及MCP相关功能时,这个Skill能提供实用的操作指南。

查看技能

web-cli-teleport

设计

该Skill帮助开发者根据任务特性选择Claude Code的Web或CLI界面,并指导如何在两种环境间无缝迁移会话。它能分析任务复杂度、迭代需求等要素,推荐最优工作界面和工作流。关键特性包括会话状态管理、环境切换指导和上下文优化建议。

查看技能