This skill guides developers through the systematic creation of a High-Performance Liquid Chromatography (HPLC) method. It helps define separation goals, select column chemistry and mobile phase, and optimize conditions like gradient and flow. Use it for analyzing non-volatile, thermally labile, or polar target analytes in solution or complex matrices.

快速安装

Claude Code

推荐
主要方式
npx skills add pjt222/agent-almanac -a claude-code
插件命令备选方式
/plugin add https://github.com/pjt222/agent-almanac
Git 克隆备选方式
git clone https://github.com/pjt222/agent-almanac.git ~/.claude/skills/develop-hplc-method

在 Claude Code 中复制并粘贴此命令以安装该技能

技能文档

Develop an HPLC Method

Systematic build of HPLC method. Mode, column chemistry, mobile phase + gradient, flow + temp, detector, iterative refinement. For non-volatile, thermally labile, or polar analytes.

When Use

  • Analyze compounds non-volatile, thermally labile, or too polar for GC
  • Build new HPLC-UV, HPLC-fluorescence, or LC-MS method from scratch
  • Adapt literature or pharmacopeial HPLC method to different column or instrument
  • Fix existing method with poor resolution, long run times, or sensitivity issues
  • Pick right chromatographic mode (reversed-phase, HILIC, ion-exchange, SEC, chiral)

Inputs

Required

  • Target analytes: Compound names, structures, MW, pKa, logP/logD
  • Sample matrix: Formulation, biological fluid, environmental extract, or neat solution
  • Performance targets: Required resolution, detection limits, quantitation range

Optional

  • Reference method: Compendial or literature method as starting point
  • Available columns: HPLC column inventory
  • Instrument config: UHPLC vs conventional HPLC, available detectors, column oven range
  • Throughput: Max run time incl re-equilibration
  • Regulatory context: ICH, USP, EPA, or other framework

Steps

Step 1: Define Separation Goals

  1. Compile analyte properties: MW, pKa, logP (or logD at relevant pH), chromophores, fluorophores, ionizable groups.
  2. Identify matrix + expected interferents (excipients, endogenous compounds, degradation products).
  3. Set perf criteria:
    • Resolution between critical pairs (Rs >= 2.0 for regulated)
    • Detection limits (LOD/LOQ)
    • Acceptable run time incl gradient re-equilibration
  4. Method purpose: assay, impurity profiling, dissolution, content uniformity, cleaning verification — drives validation category.
  5. Isocratic vs gradient: use isocratic if all analytes elute within retention factor range 2 < k' < 10; else gradient.

Got: Spec doc lists analytes + physicochemical properties, matrix, perf criteria, isocratic vs gradient decision.

If fail: pKa or logP unknown? Estimate from structure via prediction tools (ChemAxon, ACD/Labs) or run scouting gradient on C18 at pH 3 + pH 7 to empirically assess retention.

Step 2: Pick Column Chemistry

Pick chromatographic mode + column by analyte properties.

ModeColumn ChemistryMobile PhaseBest For
Reversed-phase (RP)C18 (ODS)Water/ACN or water/MeOH + acid/bufferNon-polar to moderately polar, most small molecules
RP (extended)C8, phenyl-hexyl, biphenylWater/organic + modifierShape selectivity, aromatic compounds, positional isomers
RP (polar-embedded)Amide-C18, polar-endcapped C18Water/organic, compatible with high aqueousPolar analytes that elute too early on standard C18
HILICBare silica, amide, zwitterionicHigh organic (80-95% ACN) + aqueous bufferVery polar, hydrophilic compounds (sugars, amino acids, nucleotides)
Ion-exchange (IEX)SAX or SCXBuffer with ionic strength gradientPermanently charged species, proteins, oligonucleotides
Size-exclusion (SEC)Diol-bonded silica, polymerIsocratic aqueous or organic bufferProtein aggregates, polymers, molecular weight distribution
ChiralPolysaccharide (amylose/cellulose)Normal-phase or polar organic modeEnantiomeric separations, chiral purity
  1. Default to reversed-phase C18 for small molecules with logP > 0.
  2. Analytes with logP < 0? Evaluate HILIC or ion-exchange.
  3. Particle size: sub-2 um for UHPLC (higher efficiency, higher backpressure), 3-5 um for conventional HPLC.
  4. Column dimensions: 50-150 mm length, 2.1-4.6 mm ID. Narrower columns save solvent + improve MS sensitivity.
  5. Chiral separations? Screen min 3-4 chiral stationary phases with different selectors.

Got: Column chemistry, dimensions, particle size picked with justification by analyte properties.

If fail: Initial scouting shows poor retention on C18? Switch to more retentive phase (phenyl-hexyl for aromatics) or different mode (HILIC for polar).

Step 3: Design Mobile Phase + Gradient

  1. Pick organic modifier:
    • Acetonitrile (ACN): lower viscosity, sharper peaks, better UV transparency below 210 nm
    • Methanol (MeOH): different selectivity, sometimes better for polar analytes, higher viscosity
  2. Pick aqueous component + pH:
    • Neutral analytes: water with 0.1% formic acid (MS-compatible) or phosphate buffer (UV only)
    • Ionizable analytes: buffer mobile phase 2 pH units away from analyte pKa → single ionic form
    • pH 2-3 (formic/phosphoric acid): suppresses ionization of acids, good general starting point
    • pH 6-8 (ammonium formate/acetate): for basic analytes or when low-pH selectivity insufficient
    • pH 9-11 (ammonium bicarbonate, BEH columns): for very basic compounds on high-pH-stable columns
  3. Design gradient:
    • Start at 5-10% organic, ramp to 90-95% organic over 10-20 min for initial scouting
    • Evaluate scouting chromatogram → find useful organic range
    • Narrow gradient to only elution window of interest
    • Gradient slope: steeper = faster but lower resolution; shallower = better resolution, longer run
  4. Include column wash step (95% organic, 2-3 min) + re-equilibration (initial conditions, 5-10 column volumes).
  5. Isocratic methods: target k' = 3-8 for analytes of interest.

Got: Mobile phase composition (organic, aqueous, buffer/additive, pH) + gradient profile defined. Scouting run confirms analyte elution within programmed window.

If fail: Selectivity poor (analytes co-elute despite gradient opt)? Change organic modifier (ACN to MeOH or reverse), adjust pH by 2 units, or add ion-pair reagent for charged analytes.

Step 4: Optimize Flow Rate + Temperature

  1. Set initial flow rate by column dimensions:
    • 4.6 mm ID: 1.0 mL/min
    • 3.0 mm ID: 0.4-0.6 mL/min
    • 2.1 mm ID: 0.2-0.4 mL/min
  2. Verify backpressure within instrument + column limits (usually < 400 bar conventional, < 1200 bar UHPLC).
  3. Optimize column temp:
    • Start at 30 C for reproducibility (avoid ambient fluctuations)
    • Raise to 40-60 C to cut viscosity, lower backpressure, sharpen peaks
    • Chiral columns: temp often has strong effect on enantioselectivity — screen 15-45 C
  4. Evaluate flow rate effect on resolution: small flow increases can boost throughput without big resolution loss if near van Deemter minimum.
  5. Document optimal flow rate, column temp, backpressure.

Got: Flow rate + column temp optimized. Backpressure within limits. Resolution maintained or improved vs initial conditions.

If fail: Backpressure too high? Reduce flow rate, raise temp, or switch to wider-bore or larger-particle column. Resolution degrades at higher temp? Return to 30 C, accept longer run.

Step 5: Pick Detector

DetectorPrincipleSensitivitySelectivityKey Considerations
UV (single wavelength)Absorbance at fixed lambdang rangeCompounds with chromophoresSimple, robust, most common
DAD (diode array)Full UV-Vis spectrumng rangeChromophores + spectral IDPeak purity assessment, library matching
Fluorescence (FLD)Excitation/emissionpg range (10-100x more sensitive than UV)Native fluorophores or derivatizedExcellent selectivity, requires fluorescent analytes
Refractive index (RI)Bulk propertyug rangeUniversal (no chromophore needed)Temperature-sensitive, gradient-incompatible
Evaporative light scattering (ELSD)Nebulization + light scatteringng rangeUniversal, non-volatile analytesSemi-quantitative, non-linear response
Charged aerosol (CAD)Nebulization + corona dischargeng rangeUniversal, non-volatile analytesMore uniform response than ELSD
Mass spectrometry (MS)m/z detectionpg-fg rangeStructural, highest selectivityRequires MS-compatible mobile phases
  1. UV chromophores (aromatic rings, conjugated systems)? Start with DAD — gives both quantitation + peak purity.
  2. Trace analysis, complex matrices? Prefer MS (ESI or APCI) in SIM or MRM mode.
  3. Compounds without chromophores (sugars, lipids, polymers)? Use CAD, ELSD, or RI.
  4. Set detection wavelength at analyte's absorption max (lambda-max) for best sensitivity, or 210-220 nm for general screening.
  5. Fluorescence? Optimize excitation + emission wavelengths via spectral scan of analyte.
  6. Ensure mobile phase additives compatible: no phosphate buffers with MS, no UV-absorbing additives at low wavelengths.

Got: Detector picked + configured (wavelength, gain, acquisition rate). Right for analyte chemistry + sensitivity needs.

If fail: UV sensitivity insufficient at required LOQ? Consider fluorescence derivatization (e.g., OPA for amines, FMOC for amino acids) or switch to LC-MS/MS for max sensitivity + selectivity.

Step 6: Evaluate + Refine

  1. Inject system suitability standard 6x. Evaluate:
    • Retention time RSD < 1.0%
    • Peak area RSD < 2.0%
    • Resolution of critical pair >= 2.0
    • Tailing factor 0.8-1.5 for all peaks
    • Theoretical plates per column spec
  2. Inject placebo/matrix blank to check interference at analyte retention times.
  3. Inject stressed or spiked sample to verify method separates degradation products from main analyte(s).
  4. Any criterion fails? Adjust one var at a time:
    • Poor resolution: change pH, gradient slope, or column chemistry
    • Tailing: add amine modifier (TEA for basic analytes), change buffer, or try different bonded phase
    • Sensitivity: raise injection volume, concentrate sample, or switch detector
  5. Lock final method params + document all conditions.

Got: All system suitability criteria met. Method resolves target analytes from matrix interferents + known degradation products. Params documented for transfer.

If fail: Iterative adjustment doesn't fix issue? Try fundamentally different approach (change chromatographic mode, 2D-LC, or derivatization). Return to Step 2.

Checks

  • All target analytes resolved with Rs >= 2.0 for critical pairs
  • Retention time RSD < 1.0% across 6 replicate injections
  • Peak area RSD < 2.0% across 6 replicate injections
  • Tailing factors 0.8-1.5 for all analyte peaks
  • No matrix interference at analyte retention times
  • Degradation products resolved from main analyte(s)
  • Run time (incl re-equilibration) meets throughput
  • Mobile phase compatible with selected detector
  • Method params fully documented (column, mobile phase, gradient, flow, temperature, detector)

Pitfalls

  • Ignoring mobile phase pH for ionizable analytes: Running at pH near analyte's pKa → split peaks or poor reproducibility (compound exists in two ionic forms). Buffer at least 2 pH units away from pKa.
  • Phosphate buffers with MS detection: Phosphate = non-volatile, contaminates MS source. Use formate or acetate buffers for LC-MS.
  • Insufficient re-equilibration after gradient: Column must be flushed with at least 5-10 column volumes of initial mobile phase before next injection. Inadequate re-equilibration → retention time drift.
  • Too short a column for complex mixtures: Short columns (50 mm) fast but may not give enough plates for multi-component separation. Start 100-150 mm for method dev.
  • Neglecting system dwell volume: Dwell volume (mixer to column head) delays gradient reaching column. Differs between instruments → method transfer failures. Measure + document.
  • Running HILIC like reversed-phase: HILIC needs high organic (80-95% ACN) with small aqueous fraction. Raising aqueous content → raises elution strength — opposite of RP. Equilibration times also longer.

See Also

  • develop-gc-method -- gas chromatography for volatile + semi-volatile analytes
  • interpret-chromatogram -- reading + interpreting HPLC + GC chromatograms
  • troubleshoot-separation -- diagnose + fix peak shape, retention, resolution
  • validate-analytical-method -- formal ICH Q2 validation of developed HPLC method

GitHub 仓库

pjt222/agent-almanac
路径: i18n/caveman/skills/develop-hplc-method
0
agentsagentskillsai-assisted-developmentclaude-codeskillsteams

相关推荐技能

executing-plans

设计

该Skill用于当开发者提供完整实施计划时,以受控批次方式执行代码实现。它会先审阅计划并提出疑问,然后分批次执行任务(默认每批3个任务),并在批次间暂停等待审查。关键特性包括分批次执行、内置检查点和架构师审查机制,确保复杂系统实现的可控性。

查看技能

requesting-code-review

设计

该Skill可在完成任务、实现主要功能或合并代码前自动调度代码审查子代理,确保实现符合需求和计划。它支持通过指定git SHA范围进行精准的代码变更审查,帮助开发者在关键节点及时发现潜在问题。核心原则是"早审查、勤审查",适用于开发流程的各个关键阶段。

查看技能

connect-mcp-server

设计

这个Skill指导开发者如何将MCP服务器连接到Claude Code,支持HTTP、stdio和SSE三种传输协议。它涵盖了从安装配置到认证安全的完整流程,适用于集成GitHub、Notion、数据库等外部服务。当开发者需要添加集成、配置外部工具或提及MCP相关功能时,这个Skill能提供实用的操作指南。

查看技能

web-cli-teleport

设计

该Skill帮助开发者根据任务特性选择Claude Code的Web或CLI界面,并指导如何在两种环境间无缝迁移会话。它能分析任务复杂度、迭代需求等要素,推荐最优工作界面和工作流。关键特性包括会话状态管理、环境切换指导和上下文优化建议。

查看技能